QIU Yuntao, FENG Yan, CHENG Yongjun, et al. Analysis of the Influence of RF DC/AC Ratio on the Performance of Quadrupole Mass Analyzer[J]. VACUUM AND CRYOGENICS, 2022, 28(2): 200-204. DOI: 10.3969/j.issn.1006-7086.2022.02.011
Citation: QIU Yuntao, FENG Yan, CHENG Yongjun, et al. Analysis of the Influence of RF DC/AC Ratio on the Performance of Quadrupole Mass Analyzer[J]. VACUUM AND CRYOGENICS, 2022, 28(2): 200-204. DOI: 10.3969/j.issn.1006-7086.2022.02.011

Analysis of the Influence of RF DC/AC Ratio on the Performance of Quadrupole Mass Analyzer

  • The quadrupole mass analyzer is the core part of the quadrupole mass spectrometer.The ratio of DC voltage component and high frequency AC voltage component in the RF module applied on the quadrupole will have an important influence on the performance of the quadrupole mass spectrometer.A four-order Runge-Kutta method is used to solve the differential equation,the ion movement trajectory in the quadrupole mass analyzer is simulated,and experiments are de-signed to debug the electrical parameters of the quadrupole mass spectrometer.It is found that the mass resolution of the quadrupole mass spectrometer improves and the signal intensity decreases with the increase of the ratio of RF DC to AC voltage.
  • loading

Catalog

    Turn off MathJax
    Article Contents

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return