PERFORMANCE TEXTS OF DYNAMIC FLOW METHOD UHV STANDARD APPARATUS

  • The dynamic flow method UHV standard apparatus is an absolute vacuum standard for vacuum gauge calibration.It is composed of a calibration system,a gas flowmeter and pumping system.The the ultimate vacuum is 6.5×10 -9 Pa the calibration range is from 10 -2 to 10 -7 Pa and the uncertainty amounts to 3% from 10 -2 to 10 -6 Pa and increases to 10% in 10 -7 Pa.